TC-3000 Tri-Meter:
|
|
|
Kit supplied with 0, 1, |
LaMotte Company introduces the first compact water analysis instrument designed for precision measurements of Turbidity, Chlorine and Color. The TC-3000 is equipped with an advanced microprocessor to drive a patent pending optical system that permits EPA or ISO compliant measurements for the water and wastewater analyst. The unique optics configuration enables the instrument to read low levels for Turbidity (0.05 NTU), Chlorine (0.02 ppm) and Color (0.2 PCU) while also expanding to higher ranges, 0-4,000 for Turbidity and 0-10 ppm for Chlorine. The TC-3000 is available in either EPA or ISO 7027 compliant versions and features 6 languages, data logging, RS-232 interface and CE mark. |
Turbidity Specifications: |
|
|
Unit of Measure: |
NTU, FNU, FAU, ASBC, & EBC |
|
Range: |
0-4000 |
| Resolution: | 0.01 NTU/FNU (0.00-10.99) |
| 0.1 NTU/FNU (11.00-109.9) | |
| 1 NTU/FNU (110-4000) | |
|
Accuracy: |
±2% |
|
Detection Limit: |
0.05 NTU/FNU |
|
Reproducibility: |
0.02 NTU/FNU, 0.5 FAU |
|
Light Source: |
860nm LED (ISO) Tungsten (EPA) |
Chlorine Specifications: |
|
|
Range: |
0-10 ppm |
|
Resolution: |
0.01 ppm (0-5) |
| 0.1 ppm (5-10) | |
|
Accuracy: |
0.02 or ±2% |
|
Detection Limit: |
0.02 ppm |
|
Response Time: |
<5 Seconds |
|
Light Source: |
525 nm LED |
Color Specifications: |
|
|
Range: |
0-500 CU (color units) |
|
Resolution: |
0.1 CU (0-100 CU) |
|
1 CU (100-500 CU) |
|
|
Accuracy: |
±0.5 CU or 2% |
|
Detection Limit: |
0.2 CU |
|
Response Time: |
<5 Seconds |
|
Light Source: |
UV LED 375 nm |
Meter Features: |
|
| Signal Averaging: | Disabled, 2, 5, 10 |
| Battery: | 9V |
| AC Power: | Optional |
| Data Logging: | 4000 points |
| Auto Shut-Off: | Disabled, 5, 10, 30 |
| Optional Software: | SmartLink2 |
| Response Time: | <5 Seconds |
| Languages | English, French, Spanish, Japanese, Italian, & Portuguese |
Item Description |
Order Code |
|
TC-3000e Turbidity, Chlorine, and Color Meter (turbidity complies with USEPA 180.1 Standard) |
1964-EPA |
|
TC-3000i Turbidity, Chlorine, and Color Meter (turbidity complies with ISO 7027 Standard) |
1964-ISO |
|
AC adapter (variable 100-240 VAC) |
1754 |
|
Vial Ring (2-pk) |
0641 |
|
Six pack of vials |
0475 |
|
O NTU Standard (ISO and EPA) |
1480 |
| 1 NTU Standard (ISO) | 1481 |
| 10 NTU Standard (ISO) | 1482 |
| 100 NTU Standard (ISO) | 1483 |
| 1 NTU Standard (EPA) | 1484 |
| 10 NTU Standard (EPA) | 1485 |
| 100 NTU Standard (EPA) | 1486 |
| TC-3000 Meter Manual FREE in downloadable PDF format (Just Click Here) | 1964-MN.PDF |
| TC-3000 Meter CE Compliance Certificate in downloadable PDF formate | TC-3000 CE Certificate.PDF |
| AMCO Turbidity Standards, ISO 7027 Certificate of Compliance | ISO7027.PDF |
LaMotte CompanyP.O. Box 329 | 802 Washington Avenue | Chestertown | Maryland | 21620 |